Pressure And Flowrate Regulation; Vent Exhaust - Teledyne 2010A Operating Instructions Manual

Thermal conductivity analyzer
Table of Contents

Advertisement

Thermal Conductivity Analyzer
NOTE:
NOTE:

2-4 Pressure and Flowrate Regulation

Appropriate pressure reducing regulators must be installed at all gas
supply sources. To minimize flowrate adjustments the pressure regulators on
the supporting gas supply cylinders should be adjusted to provide the same
output pressure as the sample line regulator.
The gas pressure input should be reasonably well regulated. Pressures
between .35 and 3.5 bar (5- 51 psig) are acceptable - .7 bar (10 psig) is
normal as long as the pressure, once established, will keep the flow con-
stant during analysis and within 50-200cc/min (between 0.1 and 0.4 scfh)
See Note.
Note: Gases lighter than air have a flowrate higher than indicated on
the flowmeter, while gases heavier than air have a flowrate
lower than indicated. Values can range from one half to twice
the indicated flowrate.
For example: For hydrogen or helium, set the flowrate to 0.1
scfh (50 cc/min). For carbon dioxide or argon, set the flowrate
to 0.4 scfh (200 cc/min).
When installing pressure regulators on supply cylinders, crack the
cylinder valves so that gas is flowing during installation. This will eliminate
the most common cause of standardization-gas contamination: air trapped
during assembly diffusing back into the cylinder. This procedure is particu-
larly important in applications where impurity content of 1 to 2 % is the
range of interest.

2-5 VENT Exhaust

There are two separate VENT fittings—one for the sample gas and one
for the reference gas
to minimize back pressure from restricted flow.
Exhaust connections must be consistent with the hazard level of the
constituent gases. Check local, state, and federal laws, and ensure that the
exhaust stream vents to an appropriately controlled area if required. If not
The sample-line pressure regulator should be installed as
close to the sample point as possible to minimize sample-
line lag time.
An additional option is available for SEALED reference
application. This option would not have the reference Gas
Flow Meter, Piping and Fittings.
.
1
Use
/
inch tubing for both sample and reference vents
4
Teledyne Analytical Instruments
Part II: Analysis Unit
Part II: 2-3