4.6.3 Precautions

The Model 2010A allows a great deal of flexibility in choosing ranges for
automatic range switching. However, there are some pitfalls that are to be
avoided.
Ranges that work well together are:
Range schemes that are to be avoided include:
Figure 4-2 illustrates these schemes graphically.
4.7
Normally, all of the functions automatically switch back to the Analyze
function when they have completed their assigned operations. Pressing the
Ranges that have the same lower limits but upper limits that differ
by approximately an order of magnitude
Ranges whose upper limits coincide with the lower limits of the
next higher range
Ranges where there is a gap between the upper limit of the range
and the lower limit of the next higher range.
Ranges that overlap
Ranges whose limits are entirely within the span of an adjoining
range.
Figure 4-2: Examples of Autoranging Schemes

The Analyze Function

Teledyne Analytical Instruments
Part I: Control Unit
Part I 4-21