"Firmware download
failed. Check SCSI bus."
NOTE: Firmware is
downloaded automatically
to the Scanner the first
time that you try to access
the Scanner during each
work session. If this
message appears, the
KODAK Driver is unable
to download firmware to
the Scanner.
Possible Cause(s)
The Scanner SCSI ID is
set improperly.
You have connected the
Scanner to a different
computer (or to the same
computer to which you
recently connected an
additional SCSI device),
and there is now a SCSI
ID conflict.
An unexpected problem
occurred with the
Scanner hardware.
Suggested Solution(s)
Reset the Scanner SCSI ID so that
it is different from other devices
connected to the computer. To
do so, turn off the computer then
the Scanner, set the Scanner to a
unique SCSI ID, turn on the
Scanner then the computer, and
try again.
Reset the Scanner SCSI ID so that
it is different from other devices
connected to the computer. To
do so, turn off the computer then
the Scanner, set the Scanner to a
unique SCSI ID, turn on the
Scanner then the computer, and
try again.
If the steps suggested just above
do not resolve the problem,
reinstall the Scanner software as
described in Chapter 3
(Macintosh) or Chapter 7 (PC),
and try again.