Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ Cardiovascular Imaging System Pre-Installation Manual
GE Healthcare
The loud–speaker is to be connected to the handswitch support which provide a type 3.5 mm Jack
plug for connecting external audio device.

4.3 Room Lighting

See Illustration Room Lighting for Innova System in
Drawings, for a recommended room lighting layout for an Innova LC system.
Requirements for lighting
Requirement for lighting concern the following, general, light–technique characteristics:
Illuminator level.
Lighting distribution.
Preventing the operator from being dazzled by the light (by direct light sources or by reflection
on bright objects).
The Illumination level must be compliant with established lighting technical rules and be as
constant as possible.