Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ Cardiovascular Imaging System Pre-Installation Manual
GE Healthcare
6.5 Ethernet Network Requirement
The IVUS CPU located in the control room shall be connected to the hospital Ethernet network. A
wall Ethernet outlet shall be available in the control room to connect the device.
DICOM Image Storage:
Saving patient cases to DVD: The archived images are stored in DICOM format with Volcano s5i
system acting as a File Set Creator (FSC), following the guidelines in the 2004 DICOM 3.0
specification.
Sending patient cases to DICOM server: The Volcano s5i supports the ultrasound multi-frame
image storage SOP class as an SCU (service class user).
NOTE:
All this applies to the IVUS mode of the s5i/GE, not for the FFR mode.
For more information, see the Volcano s5 DICOM Conformance Statement located in the
806365-005 Volcano Service Manual.
190
The IVUS Isolation Transformer will be connected to a standard hospital outlet.
There is no grounding requirement specified at site preparation.
Direction 5160944-13-1EN, Revision 2
6 IVUS Rev 2 Option