GE Innova 2100-IQ Preinstallation Manual page 97

Cardiovascular imaging system
Hide thumbs Also See for Innova 2100-IQ:
Table of Contents

Advertisement

Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ Cardiovascular Imaging System Pre-Installation Manual
GE Healthcare
Legend of
Illustration 4-7
1. Innova LC positioner
2. Patient table
3. C1 Cabinet
4. C2 Cabinet
5. Chiller Coolix 4000 (X–Ray tube housing)
6. VCIM Console and DL keyboard
7. DL Image monitor
8. DL Keypad
9. Flat panel monitor for Live Images
10. Flat Panel monitor for Roadmap Images
11. Chiller Innova Detector
12. XT monitor suspension
13. 21" (53 cm) FFD Monitors put on frame with monitoring screen
14. Injector on Pedestal
15. - NA -
16. Bookcase or bookshelves required for Technical Publications and special parts and tools
17. AW Station and flat panel monitor (x2)
18. Bolus handle (optional)
19. Fluoro UPS (optional)
20. Fluoro UPS I/F box (optional)
21. EMI Filter Enclosure (optional)
Chapter 4 Room Layouts
LOCATION IN TECHNICAL ROOM FOR ELECTRICAL CABINETS IS
MANDATORY.
THE ELECTRONIC CABINETS (C1, C2, COOLIX 4000 CHILLER, DETECTOR
CHILLER AND FLUORO UPS WHEN INSTALLED ) INCLUDE FANS THAT ARE
CREATING AIR–CIRCULATION OF PULSED–AIR. WHEN THIS PULSED AIR IS
IN AN ENVIRONMENT THAT MAY CONTAIN AIRBORNE PATHOGENS LIKE AN
EXAM ROOM/CONTROL ROOM, THERE IS A RISK OF TRANSMISSION OF
THESE AIRBORNE PATHOGENS FROM PATIENTS TO OTHER PATIENTS OR
CLINICAL PERSONNEL (NOSOCOMIAL DISEASES).
TO REDUCE THIS RISK, THE ELECTRONIC CABINETS MUST BE INSTALLED
IN A ROOM SEPARATED FROM EXAM ROOM/CONTROL ROOM, I.E.,
TECHNICAL ROOM.
Direction 5160944-13-1EN, Revision 2
97