Exiting Three-Point Calibration - Honeywell MiniRAE 3000+ User Manual

Handheld pid monitors
Table of Contents

Advertisement

When Span calibration is complete, you will see a message similar to this (the value shown here
is for example only):
Span 2 is done!
Reading = 1000 ppm
The instrument then exits Span calibration and shows the Zero calibration menu on its display.
Note: The reading should be very close to the span gas value.

Exiting three-point calibration

When you are done performing calibrations, press [MODE], which corresponds with "Back" on
the display. You will see the following message:
Updating settings...
The instrument updates its settings and then returns to the main display. It begins or resumes
monitoring.
Handheld PID Monitors
66
User Manual