Reverse direction navigation
Sometimes you want to go back to a previous screen rather than advance through an entire set
of screens before "wrapping around" to that screen again.
To reverse direction:
1. Press and hold [N/-] for 3 seconds.
2. When the arrow changes from pointing up to pointing down, release your finger.
Now when you press [N/-], you step back through the screens.
To change direction again: Press and hold [N/-] for 3 seconds and then release.
Note: Changing direction does not work with all screens. It works primarily in submenus.
Policy enforcement
The instrument can be configured to enforce a facility/company's requirements that calibration
and/or bump testing be performed at specified intervals, and to explicitly prompt the user that
calibration/bump testing is required. Depending on how Policy Enforcement features are
configured, the user may be required to perform a bump test or calibration prior to being able
to use the instrument. That is, it can be set to not allow normal operation of the instrument
unless calibration or bump testing is performed and passed.
If the instrument has been bump tested and calibrated in compliance with the policy settings, a
check-mark icon is included along the top of the instrument's screen:
If Policy Enforcement is enabled, then after startup the instrument displays a screen that
informs the user that the instrument requires either a bump test or a calibration. If both are
required, then they are shown in sequence.
Note: Policy enforcement features are disabled by default.