Entering calibration

1. Press and hold [MODE] and [N/-] until you see the Password screen.
2. In Basic User Level, you do not need a password to perform calibrations. Instead of
inputting a password, enter calibration by pressing [MODE].
Note: If you inadvertently press [Y/+] and change any of the numbers, simply press [MODE]
and you will be directed to the calibration menu.
The Calibration screen is now visible with Zero Calibration highlighted.
These are your options:
Press [Y/+] to select the highlighted calibration (Zero Calib or Span Calib).
l
Press [MODE] to exit calibration and return to the main display and resume measurement.
l
Press [N/-] to toggle the highlighted calibration type.
l