Calibration; Calibration Prompt; Zero Calibration (Span Calibration For Oxygen) - Honeywell X4 Series Operating Instructions And Maintenance Manual

Table of Contents

Advertisement

4. Calibration

note
Calibration should be carried out with fresh batteries.
Warning
Calibration cannot be carried out when the detector emits a low battery
alarm.
The calibration mode menu structure is shown in Appendix A (A-1 and A-).

4-1. Calibration Prompt

To enter calibration mode while in measuring mode, press the ON/OFF button 
times.
When "no" is selected, the calibration will be aborted. When "YES" is selected, the
calibration will be performed.

4-2. Zero Calibration (Span Calibration for Oxygen)

ZERO calibration must be performed in a clean atmosphere. It is recommended that
a ZERO calibration be performed daily or after any gas alarm.
The detector will initiate a ZERO calibration showing a blinking ZERO
Calibration icon and a countdown from '00' to '000'.
When the ZERO calibration has been completed successfully for all sensors,
the Test Pass icon
4
A "CAL no" prompt will be displayed.
Press the UP or DOWN buttons to alternate between "no"
or "YES" and press the ON/OFF button to select.
a
Countdown
will blink for 5 seconds.
or
X4 Series

Advertisement

Table of Contents
loading

Table of Contents