Bump status

Note: The sensor(s) should be challenged on a periodic basis.
The instrument displays this icon if it requires a bump test:
A bump test is required (and indicated by this icon) if:
The defined period between bump tests has been exceeded (bump test overdue).
l
The sensor has failed a previous bump test.
l

Glance mode

Glance Mode allows you to get vital information without turning on the instrument. You can
check information such as the instrument's model and serial number, installed sensor types,
wireless modules installed, etc., which may help when taking inventory of instruments and their
sensors or when working with service or support personnel. Glance Mode can be
enabled/disabled via Honeywell Safety Suite Device Configurator (SSDC).

Enter glance mode

Note: The instrument must be configured so that Glance Mode is turned on (the default mode
is "Off"). This can be done in Safety Suite Device Configurator (SSDC). In SSDC, Glance Mode
can be enabled or disabled by checking or unchecking the box labeled "Enable Glance Mode."
You can also set which screens will be displayed, as well as their order.
With the instrument turned off, press and hold [MODE] and [N/-] simultaneously for 2 seconds
to enter Glance Mode. If you see the message "GLANCE MODE DISABLED," you must configure
the instrument to use Glance Mode.
If Glance Mode is enabled, the first screen, with the Honeywell Logo, is displayed. Release the
[MODE] and [N/-] keys, and the first screen with information about the instrument is shown.
Note: If there is no information to show, the instrument will display "No Screen Display!" and
turn itself off.