Response time (T
):
90
Accuracy (Isobutylene):
PID Detector:
Calibration:
Inlet Probe:
Keypad:
Direct Readout:
Datalogging:
Intrinsic Safety:
EM Interference:
Alarm Setting:
Alarm:
Alarm Type:
Real-time Clock:
Communication:
Sampling Pump:
Temperature:
Humidity:
Housing (including rubber
boot):
Handheld PID Monitors
2 seconds
10 to 2000 ppm: ±5% at calibration point.
Easy access to lamp and sensor for cleaning and replacement
Two-point field calibration of zero and standard reference gases
Patented Reflex PID™ technology
Flexible 5" tubing
1 operation key and 2 programming keys; 1 flashlight switch
Instantaneous, peak value, and battery voltage
260,000 points with time stamp, serial number, user ID, site ID, etc.
ATEX
IECEx
CSAE 21UKEX2051X
Highly resistant to EMI/RFI.
Separate alarm limit settings for Low Alarm and High Alarm
Buzzer 95dB at 30cm and flashing red LEDs to indicate exceeded preset limits, low battery
voltage, or sensor failure
Latching or automatic reset
Automatic date and time
Download instrument setup from PC via RS-232 with Travel Charger, or via USB with optional
Charger Stand
Upload data to PC and download instrument setup from PC via USB on charging station.
Enhanced datalogging capability through BLE module and mobile app
Internally integrated. Flow rate: 450 to 550 cc/min.
-20º C to 50º C (-4º to 122º F)
0% to 95% relative humidity (non-condensing)
Polycarbonate, splashproof and dustproof
Battery can be changed without removing rubber boot.
2460
                                                 Ex ia IIB T4 Gb
Sira 17ATEX2082X
CSA 10.0005X Ex ia IIC T4 Gb
                             Ex ia IIB T4 Gb
EX ia IIC T4 Gb (with bat pack 059-3051-
000)
EX ia IIB T4 Gb (with bat pack 059-3052-
000)
114
II 2G Ex ia IIC T4 Gb
User Manual