Preparing The Software - Agilent Technologies E7474A Getting Started Manual

Drive test wireless solutions
Hide thumbs Also See for E7474A:
Table of Contents

Advertisement

Verifying the System
Method 3: Measuring the Noise Floor

Preparing the Software

Do This
1 Run the application program by going to the Start
2 Create a new project called:
3 Click Collection and select the Spectrum VFP
4 Click Maximize button in the upper-right corner of
5 Set the measurement software as follows:
6 In the Markers group, click Add button then To
7 In the Markers group, click Add to place a second
8 Use the mouse to drag Marker 2 to the minimum
Agilent Wireless Solutions
menu, selecting Programs, Agilent E74xx.
Verification Test 3 (Noise Floor).
(Virtual Front Panel)
the main window.
Band: Downlink
Frequency: See
Table 4-3 on page 150
values to use.
Span: 300 kHz
Averaging: Running
Averages: 100
Max and note the Marker 1 frequency and level.
marker on the display.
level of the trace. Note the Marker 2 frequency
and level.
Additional Information
The software starts.
It is recommended that you use the automatic
configuration wizard for this process. Refer to
"Automatic Configuration" on page
The display size increases to fill the available
space.
for
The level and frequency are displayed at the top-
left corner of the Spectrum VFP.
23.
149