Mask Setting - Konica Minolta CR-10 Plus Instruction Manual

Color reader
Hide thumbs Also See for CR-10 Plus:
Table of Contents

Advertisement

Various Settings/Instrument Settings

Mask Setting

— Confi gure as necessary —
"Auto Detect" is set under Mask Setting on the instrument when shipped from the factory.
When automatic detection is not possible, confi gure the settings appropriately for the
target mask attached. In such cases, automatic mask detection will be turned OFF. No
warning will be generated even if the attached mask is different from that set. Take careful
care when using.
• When "Mask Setting" is not displayed in the menu, refer to page E-59 and turn the
setting "ON" in the Instrument Menu Customization settings.
1
Go to the <MENU> screen.
MENU/OK
2
Select "Mask Setting" and confi rm.
MENU/OK
or
3
Select the type of target mask attached, and
confi rm.
MENU/OK
or
4
Go back to the <MENU> screen.
MENU/OK
Instrument Menu Customization
Ver.1.00.0000
MENU
Back
Browse Data
Language
Avg. Meas.
Tol. Judg.
Mask Setting
Auto Off
Preset
Date/Time
USB Connec.
10:00
MENU
Back
Mask Setting
Auto Detect
MAV
SAV
(Cont.)
5
Go back to the measurement screen.
MENU/OK
Memo
The mask settings have been changed, so
measurement results up to that point will be cleared.
Setting
Initial Setting
Setting
The attached target mask will be automatically
Auto Detect
detected. This is the recommended setting.
When the target mask cannot be detected
MAV
automatically, confi gure the settings appropriately for
SAV
the target mask attached.
32
10:00
MAV display
Target
87.9
L*
-0.6
a*
b *
2.9
Details