  • Page 1 SPECTROPHOTOMETER CF-300 INSTRUCTION MANUAL Before using this instrument, please read this manual.
  • Page 2 Konica Minolta, Inc. Notes on this Manual • Copying or reproduction of all or part of the contents of this manual without KONICA MINOLTA’s permission is strictly prohibited. • The contents of this manual are subject to change without prior notice.
  • Page 3: Safety Precautions

    Should liquid or metal objects AC adapter other than those specified by enter the instrument, turn the power KONICA MINOLTA is used, it may result in OFF immediately, disconnect the AC damage to the unit, fire or electric shock.
  • Page 4: Introduction

    Keep all packing materials (cardboard box, cushioning material, plastic bags, etc.) in a safe place. The CF-300 is a precision measuring instrument. They can be used to protect the instrument from impact and vibration during shipment to Konica Minolta for maintenance.
  • Page 5: Notes On Use

    Be sure to use the instrument within this range. Do not use it in areas of rapid tempera- ture changes. • Do not leave the CF-300 in direct sunlight or near sources of heat, such as stoves, etc. The internal temperature of the instrument may become much higher than the ambient temperature in such cases.
  • Page 6 In such case, the user is responsible for taking suitable countermeasures. • Make sure that the power switch is set to OFF (“ I “) when the CF-300 is not in use. • Always use the AC adapter supplied as a standard accessory (AC-A308) and connect it to an AC outlet of the rated voltage and frequency.
  • Page 7: Notes On Storage

     Notes on Cleaning • If the CF-300 becomes dirty, wipe it with a soft, clean dry cloth. Never use solvents such as thinner or benzene. • If the White Calibration Plate becomes dirty, wipe it with a soft, clean dry cloth. If dirt is difficult to remove, wipe it off with a cloth dampened with commercially-available lens cleaning solution.
  • Page 8: Table Of Contents

    Notes on Cleaning ..........5 Disposal Method ..........5 Maintenance ............5 INDEX ............6 Setting a Specimen Using the CF-300......7 (When using the benchtop base) ....24 Accessories Check ........8 Reflectance measurement ........ 24 System Diagram ........10 The sample holder for measuring the reflectance ... 25 Names and Functions of Parts ....11...
  • Page 9: Using The Cf-300

    Using the CF-300...
  • Page 10: Accessories Check

    Zero Calibration Box CM-A155 Used to perform zero calibration for measurement of reflectance. AC Adapter AC-A308 Used to supply power from an AC outlet to the CF-300. Input: 100 to 240 V 50/60 Hz 0.25-0.16 A Output: 8 V 1.5 A...
  • Page 11 Used to hold the white calibration plate and samples in standalone measurements. 1.3MP USB Camera Set CF-A32 When a USB camera is mounted on the finder of the CF-300, the measurement target can be viewed on a PC. Lens for USB Camera CS-to-C-Mount 1.3MP USB Camera...
  • Page 12: System Diagram

    Conversion Adapter CF-A31 1.3MP USB Camera CF-A29 Color Plate (14 colors) USB Cable(2m) for USB Camera IF-A39 SPECTRO PHOTOMETER CF-300 AC Adapter AC-A308 Benchtop Base CF-A26 USB Cable RS-232C Cable IF-A19 (2 m) IF-A12 (2 m) Zero Calibration Box CM-A155...
  • Page 13: Names And Functions Of Parts

    Names and Functions of Parts ⑤ ① ② ③ ④ ⑥ ⑦ ① AC adapter input socket....Connects the included AC adapter to the instrument. ② USB port ...........Connects the included USB cable to the instrument. ③ RS-232C port ........Connects the optional RS-232C cable to the instrument. ④...
  • Page 14: Measurement Procedure

    Measurement Procedure c This manual explains how to prepare the CF-300 and how to set a specimen. c The CF-300 is controlled by a PC to perform measurements. c For a description of measuring method using SpectraMagic DX (optional), refer to the SpectraMagic DX instruction manual.
  • Page 15: Connecting A Personal Computer

    (such as the optional Color Management Software SpectraMagic DX). • The USB communication port of the instrument conforms to USB 2.0. c To connect the instrument to a PC, you need to install the USB driver dedicated to the CF-300. Install the USB driver supplied with the software that enables connection and operation of the instrument.
  • Page 16: Connecting The Ac Adapter

    Failure to do so may damage the CF-300 or the AC adapter, causing a fire or electric shock. If the CF-300 will not be used for a long time, disconnect the AC adapter from the AC outlet. Accumulated dirt or water on the prongs of the AC adapter's plug may cause a fire and should be removed.
  • Page 17: Turning Power On And Off

    WARNING The CF-300 should not be operated if the CF-300 or the AC adapter is damaged, or smoke or strange odors occur. Doing so may result in a fire. In such situations, turn the power OFF immediately, disconnect the AC adapter from the AC outlet, and contact the nearest Konica Minolta-authorized service facility.
  • Page 18: Attaching A Target Mask

    Attaching a Target Mask WARNING Do not place the CF-300 on an unstable or sloping surface. Doing so may result in its falling or overturning, causing injury. Take care not to drop the CF-300 when carrying it. Be careful not to pinch hands in parts of the instrument or accessories that open and close.
  • Page 19: Calibration

    Calibration  Calibration for Reflectance Measurements After the Spectrophotometer has been switched on, zero calibration and white calibration must be performed before spectral reflectance measurements are taken. Zero calibration is performed using the Zero Calibration Box (which has a reflectance of 0%) to set the zero reference point for reflectance measurements;...
  • Page 20: Using The Zero Calibration Box

    Do not perform measurement with the specimen measuring port directed towards your face. Doing so may cause damage to your eyes . Be careful around openings in the CF-300. Failure to do so may result in fingers being trapped causing injury.
  • Page 21: Using The White Calibration Plate

    Do not perform measurement with the specimen measuring port directed towards your face. Doing so may cause damage to your eyes . Be careful around openings in the CF-300. Failure to do so may result in fingers being trapped causing injury.
  • Page 22: Setting A Specimen

    Setting a Specimen WARNING Do not use the CF-300 in places where flammable or combustible gases (gasoline fumes, etc.) are present. Doing so may cause a fire. Do not disassemble or modify the CF-300. Doing so may cause a fire or electric shock.
  • Page 23: Attaching/Detaching The Benchtop Base (Optional)

    The sample holder is positioned to ensure proper measurement before leaving the factory. For that reason, never detach the sample holder. c If the sample holder is detached by mistake, contact the nearest Konica Minolta-authorized service facility. The sample holder requires repositioning by the manufacturer.
  • Page 24: Attaching The Zero Calibration Box (When Using The Benchtop Base)

    . Pull the sample holder toward you and keep it open. . Fit the projections of the zero calibration box into the grooves on the CF-300 and Convex portion then close the sample holder to hold the box in place.
  • Page 25: Attaching The White Calibrationplate (When Using The Benchtop Base)

    Attaching the White CalibrationPlate (When using the benchtop base) CAUTION Do not perform measurement with the specimen measuring port directed towards your face. Doing so may cause damage to your eyes . Be careful not to pinch hands in parts of the instrument or accessories that open and close.Doing so may result in injury.
  • Page 26: Setting A Specimen (When Using The Benchtop Base)

    Doing so may cause a fire. Do not disassemble or modify the CF-300. Doing so may cause a fire or electric shock. The CF-300 should not be operated if it is damaged, or smoke or strange odors occur. Doing so may result in a fire.
  • Page 27: The Sample Holder For Measuring The Reflectance

     The sample holder for measuring the reflectance To measure the reflectance of film or plate-like specimens, clasp the specimen in the sample holder. When measuring specimens that cannot be clasped in the specimen holder, detach both the specimen holder and the benchtop base, and position the specimen flush against the specimen measuring port.
  • Page 28: Cleaning The Cf-300 And Accessories

    Cleaning the CF-300 and Accessories WARNING Do not disassemble or modify the CF-300 or AC adapter. Doing so may cause a fire or electric shock. The CF-300 should not be operated if it is damaged, or if smoke or strange odors occur. Doing so may result in a fire.
  • Page 29: Error Message

    The following error messages may appear when you use SpectraMagic DX CM-S100w, optional color management software, to control the instrument. If you see an error message, follow the instructions shown in the table below. If a problem persists, contact the nearest Konica Minolta-authorized service facility. Event Displayed message...
  • Page 30 Something is wrong with the Turn power to the instrument instrument’s memory. OFF, then back ON. If this error message does not clear, contact the nearest Konica Minolta- authorized service facility. Light receiving The light receiving circuit is Turn power to the instrument circuit error not working properly.
  • Page 31: Troubleshooting Guide

    TROUBLESHOOTING GUIDE If a problem occurs with the Spectrophotometer, please check the following points before requesting service. If the problem continues to occur even after the suggested corrective actions have been taken, contact the nearest Konica Minolta-authorized service facility. Refer Condition...
  • Page 33: Explanations

    Explanations...
  • Page 34: Illumination/Viewing System

    The flow of measurement is shown below. The geometry of the CF-300 conforms to ISO 7724/1 and JIS Z 8722 condition c (diffused illumination/perpendicular viewing system) standards, and offers both di:8° (SCI: specular component included ; Total reflectance) and de:8°...
  • Page 35: Dimensions

    Dimensions (Unit: mm) 316.6 321.4 (NC Mask), 322.4 (MAV Mask, XUSAV Mask) Ø5 hole Measurement point Contact 5 x 7 6-M6 measurement MAV Mask, XUSAV Mask long hole Tripod socket Ø5 hole Measurement point NC Mask Measurement point Noncontact 5 x 7 6-M6 measurement NC Mask...
  • Page 36: Specifications

    Within E*ab 0.15 Inter-instrument (Based on MAV/SCI contact measurements of 12 BCRA Series II color tiles compared agreement to values measured with a master body under Konica Minolta standard conditions) Interface USB 2.0, RS-232C Power Dedicated AC adapter (AC 100 to 240 V, 50/60 Hz)
  • Page 37 < CAUTION > KONICA MINOLTA WILL NOT BE LIABLE FOR ANY DAMAGES RESULTING FROM THE MISUSE, MISHANDLING, UNAUTHORIZED MODIFICATION, ETC. OF THIS PRODUCT, OR FOR ANY INDIRECT OR INCIDENTAL DAMAGES (INCLUDING BUT NOT LIMITED TO LOSS OF BUSINESS PROFITS, INTERRUPTION OF BUSINESS, ETC.) DUE TO THE USE OF OR INABILITY TO USE THIS PRODUCT.