Various Settings/Instrument Settings

Avg. Meas.

This section explains how to perform measurement when average measurement has
been confi gured. The average times icon and the number of measurements will be
displayed on the screen.
1
With the target to be used displayed on the
instrument's LCD screen, set the instrument's
sample surface aperture on the sample to
measure.
2
Measurement will be repeated until the set
number of measurements is reached.
Memo
• During averaging measurement, the current
measurement count will be displayed.
• The buzzer will emit a "beep" at every measurement,
and the color difference average will be displayed on
the LCD screen until the number of measurements is
reached.
• To continue measuring, wait at least 2 seconds after
measuring before continuing.
• To correct data during measurement, perform the
following procedure.
MENU/OK
or
 Undo 
 (return to measurement screen)
Memo
• Repeat this procedure to correct multiple data
entries.
• The amount of the cleared count will be subtracted
from the measurement count display.
10:00
05
Target
87.9
L*
-0.6
a*
b *
2.9
MENU/OK
(Cont.)
3
After the number of averaging measurements
is reached, the buzzer will emit a "beep" and
the color difference average will be displayed.
Memo
• When the color difference exceeds the tolerance, the
buzzer will emit a "beep beep".
In addition, items that have exceeded the tolerance
will be highlighted.
10:00
2.0
05
05/05
E*
6.3
L* -6.3
a*
0.1
b* -0.6
Notes
After an average is determined for the instrument, because the average is
converted for each color space and displayed, the simple averages for
measurements of each count may not match.