oCut a piece of Packing(A)[thickness 3mm] into a strip of 165mm long, and attach it to the Upper
Light shield, and fix it to the light shield panel(right and left) using the speed nuts 4 pieces of M4 x 6
pan screws from the light room side.
Pan-head Screw
(M4 x 6)
!0 C ut pieces of Packing(D)[thickness 10mm] into two strips of 20mm long and attach them to the left
and right corners of Upper Light Shield as shown in the diagram below(left).
!1 A ttach the triangle-shaped Packing(E) to the corners as shown in the diagram below(right).
Upper Light Shield
20mm Strip of Packing(D)
(One for each corner)
Pan-head Screw
(M4 x 6)
165mm Strip of
Packing(A)
Left Light Shield
Packing(E)
(One for each corner)
Main Unit Stand
INSTALLATION PROCEDURE
Upper Light Shield Panel
165mm Strip of
Packing(A)
Right Light Shield
Right Light
Shield
Lower Light Shield
SERVICE MANUAL
93