Konica Minolta SRX-101A Service Manual page 105

Medical film processor
Hide thumbs Also See for SRX-101A:
Table of Contents

Advertisement

INSTALLATION PROCEDURE
I
Adjusting Procedure
1) Checking of the CPU board (Fig. B)
Check the Jp1 on the CPU board if its pins 2-3 are selected.
2) Start of SRX-101A
Turn on the power switch, set the DEV temp. control knob at 34:, and turn on the RUN switch.
3) Actual temperature checking (Fig. F)
qWait 5 minutes after READY lamp ON, then open the top cover.
wMeasure the actual DEV temperature with a thermometer.
eFurther adjustment is not necessary if the temperature is between 33: and 35:.
SERVICE MANUAL
100
JP1-1, 2 are in short circuited.
Fig. B
Fig. F