No.
TROUBLE
6
Discoloration
• The fixer solution temperature
is too low.
• The wash water temperature
is too low.
• The fixer is exhausted due to
overuse. (when processing
small film lots.)
• There is no water in the
WASH Water tank.
7
Insufficient drying • Incorrect temperature setting.
The drying heater (H1) is
faulty.
The drying heater fans (FM1,2)
are faulty.
The drying sensor (RB1)
is faulty.
8
Uneven drying
• Incorrect temperature setting.
The drying heater (H1) is
faulty.
The drying heater fans (FM1,2)
are faulty.
The drying sensor (RB1) is
faulty.
• The wipe-down roller is dirty.
CAUSE
• Replace the fixer
solution.
• Replace the faulty
parts.
• Correct the temperature
setting.
• Replace the faulty parts.
• Correct the temperature
setting.
• Replace the faulty parts.
• Clean or replace
the wipe-down roller.
TROUBLE SHOOTING
RESPONSE
• Show the proper method for
maintaining optimum fixer temperature.
• Install a water heating device.
• Remind about processing solution
replacement intervals.
• Show checking method of the water
supply valve and wash water drainage
valve.
ADVICE To CUSTOMER