INSTALLATION INSTRUCTIONS AND PRECAUTIONS
13-15. Removing Trapped Air from the Supply Hoses
q
Turn on the RUN Lamp and keep pushing READY button to activate the auto-fill function until
the READY lamp's flashing pattern changes.
w
Lift the supply hose up so that any trapped air will move towards the processor.
Refer to the following diagram.
e
Gradually lift the hose to move the trapped air towards the processor, the trapped air will
eventually be released into the tanks.
r
After the trapped air is released, push RUN button OFF and reset the solution replenish
mode.
Trapped Air
Replenisher Tank
SERVICE MANUAL
80
Move the air towards
the processor
Processor
Support (Stand)
Supply Hose
Floor