13-6. Room Conditions Around the SRX-101A
q
The acceptable room conditions for operating the SRX-101A are :
Temperatures between 15: and 30:(59<and 86<)
Humidity levels between 30%RH and 75%RH
w
In order to maintain the above room conditions, the facility where the SRX-101A is located
should be equipped with a ceiling mounted ventilation fan and a ventilation opening in the
wall below the height of the processor.
Ex. Darkroom Ventilation
Ventilation Fan
The ventilation wall opening should be as close to the processor as possible and as far
away from the ventilation fan as possible. Use light shielding materials for both the fan and
wall openings.
e
Keep all fluorescent materials and combustible materials (thinner, benzene, gasoline.etc)
away from the processor. To prevent fires, store such dangerous materials in another
room.
r
Divert heat from burners and space heaters away from the processor. Divert forced air
from air conditioning equipment away. Being exposed to direct blast of hot or cold air, poor
film processing quality may result.
INSTALLATION INSTRUCTIONS AND PRECAUTIONS
Darkroom
SRX–101A
Ventilation Opening
(below the height of the processor)
SERVICE MANUAL