17. The READY lamp will
not come ON.
Has 15 minutes passed since
the RUN lamp came ON?
YES
Has initial wash water
supply been completed
(LED3 OFF)?
YES
Is the film entry detector ON
(The film insertion alarm will
sound).
NO
Is 0.5V or less being supplied to
JJ2 #4 and #7?
YES
Replace the Control Panel
board.
18. The READY lamp will not
turn OFF (RUN lamp)
Has 1 minute passed since
all films being processed
and exitted?
YES
Replace the CPU board.
SERVICE MANUAL
37
NO
Wait the required time.
NO
Wait the required time.
(about 2 minutes)
Check if there is any film
YES
jammed at the detector unit
or not.
NO
Replace the CPU board.
NO
Wait at least 1 minute, and
then press the RUN switch.
If all of the checks are OK,
replace the film entry
detector unit.