If the WebLink startup screen does not appear, it means that communications are not
established between the code reader and the PC. Please check the following.
• Does the code reader and the PC have a proper physical (cable) connection?
• Are the IP Addresses of the PC and code reader set correctly?
→ Refer to step 4 for setting the IP address of the PC.
• Do a hardware reset of the code reader.
→ When turning ON the power supply, press and hold the setup button on the code
reader body until its light turns on.
For other measures that can be taken, please refer to When unable to access by WebLink
in Q&A in Appendices of the MicroHAWK V320-F/V330-F/V420-F/V430-F Series Barcode
Reader User Manual (Cat. No. Z432).
8
The WebLink screen appears.
9
Click on the Setup tab and, in
Read Cycle Sequence, set
Cycle to Triggered.
10
Click on the gear icon on the
upper right of the screen to
select Advanced.
7. EtherNet/IP Connection Procedure
21