VIDEOMARK?
(IEEE-488, RS-232)
Returns the VIDEO line marker location.
Parameters: None.
Response:
*WAI
(IEEE-488, RS-232)
(Wait-to-Continue command) Prevents further remote commands from being executed
until all previous remote commands have been executed. For example, if you send an
OUT command, you can cause the 5820A Calibrator to wait until the output has settled
before continuing on to the next command if you follow OUT with a *WAI command.
The *WAI command is useful with any overlapped command, preventing the 5820A
Calibrator from processing other commands until the overlapped command is processed.
Example:
Process all existing commands before continuing.
WAVE
(IEEE-488, RS-232, Overlapped, Coupled)
(Waveform command) Sets the output waveforms for WAVEGEN mode. Waveform
choices are SINE (sinewave), TRI (trianglewave), SQUARE (squarewave), or NONE
(waveform does not apply).
Parameter:
Example:
WAVE?
(IEEE-488, RS-232)
(Waveform query) Returns the output waveform. Waveform choices are SINE
(sinewave), TRI (trianglewave), SQUARE (squarewave), or NONE (waveform does not
apply).
Responses:
Example:
Return SQUARE when the WAVEGEN ac primary output is a squarewave.
ZERO_MEAS
(IEEE-488, RS-232, Overlapped)
(in MEASV mode, TZCAP range)
Sets the measurement offset to the capacitance value to the most recent measurement if
ON, to zero if OFF.
Parameters: (boolean) ON or OFF.
6-46
(number) Line marker location.
*WAI
<waveform> , (SINE, TRI, SQUARE, NONE)
WAVE SINE
< waveform> ,
(SINE, TRI, SQUARE, NONE)
WAVE? returns SQUARE