Fluke 5820A Operator's Manual page 169

Oscilloscope calibrator
Hide thumbs Also See for 5820A:
Table of Contents

Advertisement

OUT?
(IEEE-488, RS-232)
Returns the output amplitudes and frequency of the 5820A Calibrator. Multipliers (e.g., K
or M) are not used in the response.
Response:
<primary amplitude value>,<primary units>,
<secondary amplitude value>,<secondary units>,
<fundamental frequency value>
Examples:
-1.520000E+01,V,+0.000000E+00,00.000E+00 (-15.2V)
+1.883000E-01,A,+0.000000E+00,04.420E+02 (188.3 mA, 442 Hz)
OUT_ERR?
(IEEE-488, RS-232)
(Output Error query) Returns the UUT error and units computed by the 5820A Calibrator
after shifting the output with the INCR command. The UUT error is computed only when
editing VOLTAGE mode amplitude and MARKER period.
Response:
1. (Float)
2. (Character) Units for the above number (PPM, PCT, DB, or 0)
Example:
OUT_ERR? returns -1.00000E+01,PCT
Return -10 % when the UUT is reading low by 10 %.
OUT_IMP
(IEEE-488, RS-232, Overlapped, Coupled)
Programs the oscilloscope's output impedance. The scope output impedance is selectable
only in the volt scope mode.
Parameters: Z50
Z1M
Example:
OUT_IMP Z50
UUT error magnitude
(Programs the oscilloscope's output impedance to 50 Ω.)
(Programs the oscilloscope's output impedance to 1 MΩ.)