CAL_ABORT
(IEEE-488, RS-232, Overlapped)
(Calibration Abort query) Instructs 5820A to abort calibration procedure after present
step.
Example:
CAL_ABORT
CAL_BACKUP
(IEEE-488, RS-232, Overlapped)
Skips to the next entry point in the calibration procedure.
CAL_CONST?
(IEEE-488, RS-232)
(Calibration Constant command/query) Retrieves the value in use of the given calibration
constant.
Parameter:
The name of the calibration constant
Response:
(float) The value of the constant
Example:
CAL_CONST? VDAC_G
CAL_DATE?
(IEEE-488, RS-232)
(Calibration Date command) Returns date associated with active calibration constants as
entered on the front panel or entered with the CAL_STORE command.
Parameter:
MAIN, OPT5, and GHZ
Example:
CAL_DATE? MAIN
CAL_FACT
(IEEE-488, RS-232)
(Calibration Fact command) Sets the procedure "fault action" flag.
Parameter:
(char) CONT to continue on faults or ABORT to abort on faults)
Example:
CAL_FACT ABORT (this is the default)
CAL_FACT?
(IEEE-488, RS-232)
(Calibration Fact query) Gets the procedure "fault action" flag.
Response:
(char) CONT or ABORT or BACKUP (BACKUP backs up to previous
entry)
Example:
ABORT