ISCE1
(IEEE-488, RS-232)
(Instrument Status 1 to 0 Change Enable command) Loads the two bytes into the 16-bit
ISCE1 register. (See Chapter 5 for more information.)
Parameter:
<value> (decimal equivalent of the 16 bits, 0 to 32767)
Example:
ISCE1 4108
Load decimal 4108 (binary 0001000000001100) to enable bit 12 (SETTLED).
ISCE1?
(IEEE-488, RS-232)
(Instrument Status 1 to 0 Change Enable query) Returns the two bytes from the 16-bit
ISCE1 register. (See Chapter 5 for more information.)
Response:
<value> (decimal equivalent of the 16 bits, 0 to 32767)
Example:
ISCE1? returns 4108
Return decimal 4108 (binary 0001000000001100) if bit 12 (SETTLED)is set to 1.
ISCR?
(IEEE-488, RS-232)
(Instrument Status Change Register query) Returns and clears the contents of the
Instrument Status 0 to 1 Change Register (ISCR0) and Instrument Status 1 to 0 Change
Register (ISCR1). (See Chapter 5 for more information.)
Response:
<value> (decimal equivalent of the 16 bits, 0 to 32767)
Example:
ISCR? returns 4108
Return decimal 4108 (binary 0001000000001100) if bit 12 (SETTLED) is enabled (set to
1) in the Instrument Status 0 to 1 Change Register and bit 12 (SETTLED) is enabled (set
to 1) in the Instrument Status 1 to 0 Change Register.
ISCR0?
(IEEE-488, RS-232)
(Instrument Status 0 to 1 Change Register query) Returns and clears the contents of the
Instrument Status 0 to 1 Change Register.
Response:
<value> (decimal equivalent of the 16 bits, 0 to 32767)
Example:
ISCRO? returns 4108
Return decimal 4108 (binary 0001000000001100) if bit 12 (SETTLED) is set to 1.
ISCR1?
(IEEE-488, RS-232)
(Instrument Status 1 to 0 Change Register query) Returns and clears the contents of the
Instrument Status 1 to 0 Change Register.
Remote Codes
Summary of Commands and Queries
6
6-21