Fluke 5820A Operator's Manual page 184

Oscilloscope calibrator
Hide thumbs Also See for 5820A:
Table of Contents

Advertisement

5820A
Operators Manual
UUT_FLUSH
(IEEE-488, RS-232)
(Flush UUT Receive Buffer command) Flushes the UUT receive buffer for data received
from the UUT over the 5820A Calibrator rear panel SERIAL 2 TO UUT serial port. The
command may be sent over gpib or RS-232 ports, but applies to SERIAL 2 TO UUT
serial port operation.
Parameter:
Example:
Flush the 5820A Calibrator receive data buffer for the UUT.
UUT_RECV?
(IEEE-488, RS-232)
(Return UUT Receive Data command) Returns data from the UUT serial port. The
command may be sent over gpib or RS-232 ports, but applies to SERIAL 2 TO UUT
serial port operation.
Response:
Example:
Return (for example) a measurement from the UUT. The format is #2 (two numbers
follow) 11 (characters follow) +1.99975E+0 (11 characters).
UUT_RECVB?
(IEEE-488, RS-232)
Returns binary data as integers from the UUT serial port.
Parameter:
Response:
Example:
6-42
(None)
UUT_FLUSH
(String) Data from the UUT serial port.
UUT_RECV? returns #211+1.99975E+0
(Integer) Optional maximum number of integers per line
(Indefinite ASCII) Comma separated integers as follows:
1.
(Integer) Number of data bytes being returned not including the
count
2-n. (Integer) Data from the UUT serial port as integers.
In a Fluke 45 is hooked up to the UUT port and it returns => followed by
a carriage return and a line feed, UUT_RECVB will return 4, 61, 62, 13,
10.