Leveled Sinewave Verification: Flatness - Fluke 5820A Operator's Manual

Oscilloscope calibrator
Hide thumbs Also See for 5820A:
Table of Contents

Advertisement

7-16. Leveled Sinewave Verification: Flatness

Nominal Value
(V p-p)
Frequency
0.005
50 kHz
0.005
1 MHz
0.005
10 MHz
0.005
50 MHz
0.005
100 MHz
0.005
150 MHz
0.005
200 MHz
0.005
250 MHz
0.005
300 MHz
0.005
350 MHz
0.005
400 MHz
0.005
450 MHz
0.005
500 MHz
0.005
550 MHz
0.005
600 MHz
0.0075
50 kHz
0.0075
1 MHz
0.0075
10 MHz
0.0075
50 MHz
0.0075
100 MHz
0.0075
150 MHz
0.0075
200 MHz
0.0075
250 MHz
0.0075
300 MHz
0.0075
350 MHz
0.0075
400 MHz
0.0075
450 MHz
0.0075
500 MHz
0.0075
550 MHz
0.0075
600 MHz
0.0099
50 kHz
0.0099
1 MHz
0.0099
10 MHz
0.0099
50 MHz
0.0099
100 MHz
0.0099
150 MHz
0.0099
200 MHz
0.0099
250 MHz
0.0099
300 MHz
0.0099
350 MHz
0.0099
400 MHz
0.0099
450 MHz
0.0099
500 MHz
0.0099
550 MHz
0.0099
600 MHz
Table 7-12. Leveled Sinewave Verification: Flatness
Measured
Value (V p-p)
Deviation
1-Year Spec.
(V p-p)
na
na
na
Maintenance
Verification Tables
(V p-p)
na
0.000175
0.000175
0.000175
0.000175
0.0002
0.0002
0.0002
0.0002
0.0003
0.0003
0.0003
0.0003
0.0003
0.0003
na
0.0002125
0.0002125
0.0002125
0.0002125
0.00025
0.00025
0.00025
0.00025
0.0004
0.0004
0.0004
0.0004
0.0004
0.0004
na
0.0002485
0.0002485
0.0002485
0.0002485
0.000298
0.000298
0.000298
0.000298
0.000496
0.000496
0.000496
0.000496
0.000496
0.000496
7