Fluke 5820A Operator's Manual page 104

Oscilloscope calibrator
Hide thumbs Also See for 5820A:
Table of Contents

Advertisement

5820A
Operators Manual
8. Verify the 5820A Calibrator is powered and in the reset condition. (If in doubt, press
9. In the Main Control Window, select the Remote command from the Control menu.
10. Click OK. Observe the 5820A Calibrator Control Display changes to REMOTE
11. In a similar manner, select the Local command from the Control menu, then click OK
12. In the Main Control Window, select the End GPIB Test command from the Exit
5-10
the o key on the 5820A Calibrator front panel.)
This brings the focus back to the Parameter Input Window with the only selectable
item the Listeners or Device List check boxes. Check box 4, or the corresponding
GPIB Port Address you selected for the 5820A Calibrator. (Below is typical.)
CONTROL (below).
f
f
If the Control Display does not change to REMOTE CONTROL, the IEEE-488
interface is not working properly. Check Step 3 of the "IEEE-488 Port Setup
Procedure" to verify gpib is selected as the Host port, and to verify the gpib port
address used in this procedure. If all is correct, check the ZT-1444A jumper and DIP
switch settings (Appendix C), and finally cabling between the PC and 5820A.
in the Parameter Input Window. Observe the 5820A Calibrator Control Display
changes back to the reset condition (below).
CHANGE
SCOPE Z
to DC
1 MΩ
f
f
menu.
REMOTE CONTROL
Go to Local
f
f
V/DIV
TRIG
MENU
off
f
f
oq021f.bmp
f
MODE
volt
f