4-42. Testing Video Triggers

f
Mark =
New mark =
f
To access the VIDEO menu, press ã and then select VIDEO or press the softkey
under MODE until "video" appears.
The following describes each option in the VIDEO menu:
LINE MK Allows you to select the marker line number. For ntsc and pal-m formats,
you can also select field ("odd" or "even"). For pal and secam formats, the field
("ODD" or "EVEN") is selected automatically based on marker line number.
FORMAT Scrolls through the available formats. You can select ntsc, pal, pal-m,
and secam.
MODE Indicates the calibrator is in "video" mode. Use the softkey to change modes
and open menus for other oscilloscope calibration modes.
Default video settings are + 100 %, format = NTSC, and videomark = 10.
LINE MK
10 ODD
f
f
|
|
|
|
|
|
|
|
|
|
|
10
f
f
Front Panel Operation

Testing Video Triggers

FORMAT
MODE
ntsc
video
f
f
ntsc
volt
pal
edge
pal-m
levsine
secam
marker
wavegen
video
overld
meas Z
pulse
auxin
FIELD
odd
f
f
odd
even