Verification standards supported

The optical readers without a verification license support the following verification standards:
• Uncalibrated verification according to ISO/IEC 29158 and ISO/IEC 16022 for data matrix
• Uncalibrated verification according to ISO/IEC 15416 for various 1D bar codes
The optical readers with a "Veri-Genius" license also support the following verification
standards:
• Data matrix verification to ISO/IEC 15415:2004
• Data matrix verification to AS9132 Rev A:2005 (previously IAQG) for dot peen markings
• Siemens DPM
• Data matrix verification to ISO/IEC 29158:2011
• Barcode verification to ISO/IEC 15416:2000 (previously ANSI X3.182-1990)
Note
Veri-Genius license
For productive use, you require a Veri-Genius license (article number "6GF3400-0SL02").
SIMATIC MV500
Operating Instructions, 03/2021, C79000-G8976-C494-05
Incorrect overall geometry of the marking
Damaged surface of the marking or part
Very little or non-uniform contrast in the marking
Image processing
3.2 Code verification