PS200
PORTABLE GAS MONITOR
5. Warnings & Faults
5.1. LOW BATTERY
LOW BATTERY flashes when approximately 30 minutes of run time remains, as shown in
Figure
29: LOW BATTERY
Warning. The display turns red, the audible alarm sounds once every two
seconds, and the red LED's flash.
NOTE: Gas alarms still operate if a LOW BATTERY warning exists.
Figure 29: LOW BATTERY Warning
5.2. BAT FAULT
BAT FAULT flashes when approximately 3 minutes of run time remains, as illustrated in
Figure 30:
BAT FAULT
Warning. The display turns red, the audible alarm sounds continuously and the red
LED's remain on. After 3 minutes, the monitor automatically switches OFF.
Figure 30: BAT FAULT Warning
5.3. Zero Fault
5.3.1. Zero Fault – At The End Of Warm-up
If the monitor is switched on in gas and a sensor fails to zero correctly, the display turns red and
the audible/visual alarms activate. A wrench symbol
, alternating with a gas reading, will
indicate the faulty sensor, as shown in
Figure 31: LEL ZERO
Fail.