Commissioning

6.3 Language and illumination

6.2

Write protection

The local display can be write-protected to prevent unauthorized manipulation by entering a
freely selectable, personal code number.
In this case the device functions and parameters can only be changed after entering this code.
You define the personal code number in the "6.2 Customer code" menu item.
If you actuate the
prompts you to enter the code. You can also enter the code number in the "6.1 Enter Code" menu
item. The programming is disabled again:
• after returning to display mode
• about 10 minutes after actuating an operating element
• after entering any number not the same as the code number in the "6.1 Enter Code" menu
item
Note
With code = 0 (factory setting), the programming is always enabled.
You can always disable the operation by HART communication.
6.3
Language and illumination
The desired operating language can be set in the control and display unit in menu item "1.1.1
Language".
The background illumination of the LCD is switched on automatically the first time an operating
element is actuated. The illumination ceases about 10 minutes after last actuating an operating
element. If the display is constantly lit, you can set this using the menu item "1.1.5 Illumination".
44
and
operating elements in the parameter setting level, the display
TRANSMAG 2 with MAG 911/E
Operating Instructions, 03/2021, A5E00102775-AC