Siemens SITRANS F TRANSMAG 2 Operating Instructions Manual page 86

Electromagnetic flometers
Table of Contents

Advertisement

Technical specifications
Outputs and inputs
Outputs and Inputs
Electrical isolation
Current output
Signal range
Upper limit
Failure signal
Load
Output
Communication
Protocol
Digital output 1
Signal
Output
Active signal
Passive signal
Output configuration
Pulse
Significance
Pulse width
Frequency
Limits
Digital output 2 (relay)
Relay
Rating
Output configuration
Digital input
Input function configurable high or low
active
Signal voltage
High level
Low level
PROFIBUS-PA
Communication
Transmission
Protocol layer
Device class
86
Outputs electrically isolated from each other and from the
power supply, max. 60 V permissible against PE/ equipoten‐
tial bonding
0/4 ... 20 mA
(Only 20 mA / HART devices (7ME5034-0.... and -2.....))
0/4 ... 20 mA, selectable
20 ... 22.5 mA
optional: 3.6 mA, 20 mA or 24 mA
Max. 600 Ω, load voltage max. 15 V
For HART communication: ≧ 250 Ω
Via analog output with PC coupling module or HART com‐
municator
HART version 5.1
Configurable for positive or negative logic
DC 24 V, ≦ 24 mA, Ri = 170 Ω
Open collector, max. DC 30 V, 200 mA
≦ 5000 pulses/s
≧ 0.1 ms
≦ 10.000 Hz
For flow and quantity, flow direction, alarm
only 7ME5034-0....
NC or NO function
Rating max. 5 W, UC 50 V, 200 mA
limits for flow and quantity, flow direction, alarm
Only as alternative to digital output 2 (only 7ME5034-2....)
Set measured value or counter to zero
max. 30 V, Re = 3 kΩ
DC +11 to 30 V
DC -30 to +5 V
For PROFIBUS devices (7ME5034-1....)
Layer 1 and 2 according to PROFIBUS-PA
According to IEC 1158-2
Layer 7 according to PROFIBUS-PA and DP V1 (EN 50 170)
B
Operating Instructions, 03/2021, A5E00102775-AC
TRANSMAG 2 with MAG 911/E