Siemens SITRANS F TRANSMAG 2 Operating Instructions Manual page 25

Electromagnetic flometers
Table of Contents

Advertisement

• Use proper gaskets according to liner type.
• If earthing rings are used, use standard flange gaskets on both sides of the ring.
Figure 4-10
Figure 4-11
Vibrations
In installations with strong vibrations remote installation of the transmitter is recommended.
Alternatively mitigate vibrations by installing pipe support in close proximity to the flowmeter.
Figure 4-12
TRANSMAG 2 with MAG 911/E
Operating Instructions, 03/2021, A5E00102775-AC
Correct installation with gaskets
Inlet protection
Avoid vibrations
Installing/mounting
4.2 Sensor installation
25