Digital input (Menu item 4.3)
The digital input is available with the device variant 7ME5034-2xxxx (Hart communication and
digital input). You have the following setting options in this menu:
• Function, menu item 4.3.1
By applying an external power supply, you can create a switching input, for example using a
switch (see also Figure 5-12). You have the following possibilities depending on the
configuration:
– Set the measured value to "0" (all outputs and displays are reset to zero and the totalizers
– Reset the forwards totalizer
– Reset the reverse totalizer
– Reset the net totalizer
– Reset all totalizers
• Signal type, menu item 4.3.2
You can determine the signal type (high-active) with this.
8.6

Function Group Identification

Funct. unit (Menu item 5.1)
You can call or enter device-specific or tag-related data in this menu.
You have the following options:
• TAG (tag number), menu item 5.1.1
• Tag description, menu item 5.1.2
• Message, menu item 5.1.3
TRANSMAG 2 with MAG 911/E
Operating Instructions, 03/2021, A5E00102775-AC
are stopped
elektronische
Sicherung
5 V
Ri 3k
Electronic fuse
External switch
External supply 11 ... 24 V (max. 30 V)
Figure 8-9
Circuit diagram of digital input
8.6 Function Group Identification
ext. Schalter
+
3
4
-
Functions
ext. Versorgung
max. 30 V
71