Iwatsu SY-8218 Instruction Manual page 65

Table of Contents

Advertisement

Displayed messages
W63: C drive cannot save!
W65: Target equipment is not found!
W66: Read only file cannot be saved!
W68: Already exits. Do you want
to overwrite it?
W69: Failed in a saved!
W70: Failed in a recalled!
W71: DCS output time out!
i01: B Coil has been changed.
Please enter the N2!
i02: B Coil has been changed.
Press the EXIT key to start!
Table 7-1 Error/ warning messages
Causes
 Data or setup tries to be saved in C drive.
 No storage equipment to save / recall is found
out (removed on the way).
 Recalled the data of a different model.
 Data cannot be saved in the read-only file.
 Data tries to be overwritten on an existing file.
 Data fails to be saved. (Memory is full or
memory is removed during save operation.)
 It failed in the recall of data. (SST that doesn't
correspond to the main unit is connected and so
on.)
 The electric current output time of SY-961(DCS)
exceeded the time limit.
 Because B coil was changed to User by the
Single Seat Measurement, it is urged to input
N2.
 It is informed that the kind of B coil of the
attachment was changed by the Single Seat
Measurement.
 Connect USB memory to this product and save
data in it.
This product does not disclose C drive to a
user.
 Press the EXT key and then insert USB memory
to save / recall data.
 The data of a different model can not be recalled.
 Specify other than read-only file for a storage file.
 Select Yes or No.
 Remove the cause.
 Remove the cause.
 Try to increase Tolerance.
 Decrease the number of Retry times.
 Input the value of N2.
 Press the EXIT key to continue the measurement.