Displayed messages
W63: C drive cannot save!
W65: Target equipment is not found!
W66: Read only file cannot be saved!
W68: Already exits. Do you want
to overwrite it?
W69: Failed in a saved!
W70: Failed in a recalled!
W71: DCS output time out!
i01: B Coil has been changed.
Please enter the N2!
i02: B Coil has been changed.
Press the EXIT key to start!
Table 7-1 Error/ warning messages
Causes
Data or setup tries to be saved in C drive.
No storage equipment to save / recall is found
out (removed on the way).
Recalled the data of a different model.
Data cannot be saved in the read-only file.
Data tries to be overwritten on an existing file.
Data fails to be saved. (Memory is full or
memory is removed during save operation.)
It failed in the recall of data. (SST that doesn't
correspond to the main unit is connected and so
on.)
The electric current output time of SY-961(DCS)
exceeded the time limit.
Because B coil was changed to User by the
Single Seat Measurement, it is urged to input
N2.
It is informed that the kind of B coil of the
attachment was changed by the Single Seat
Measurement.
Connect USB memory to this product and save
data in it.
This product does not disclose C drive to a
user.
Press the EXT key and then insert USB memory
to save / recall data.
The data of a different model can not be recalled.
Specify other than read-only file for a storage file.
Select Yes or No.
Remove the cause.
Remove the cause.
Try to increase Tolerance.
Decrease the number of Retry times.
Input the value of N2.
Press the EXIT key to continue the measurement.
53
B-H ANALYZER SY-8218 / SY-8219
Remedies