Displayed messages
W19: H range over at 1st excitation!
W20: B range over at 1st excitation!
W23: SST serial communication
is abnormal!
W24: DCT serial communication
is abnormal!
W25:DCS serial communication
is abnormal!
W30: Setting condition error! V2
200V
W32: Setting condition error! I1 6A
W34: Setting condition error!
Yoke V3 or V4 > 200V
W35: Setting condition error!
⊿VL > 400V
W36: Setting condition error!
⊿IL > 12A
W41: OSC output under! 0V
W42: Shortage of the exciting current!
W43: OSC output under! ≦ 0V
W50: Measured V2m over!
> 200V
Table 7-1 Error/ warning messages
Causes
Overflow occurs in the exciting current
measurement range during approximate value
excitation.
Overflow occurs in the induced voltage
measurement range during approximate value
excitation.
Serial communication with SY-956 (SST) is
abnormal.
Serial communication with SY-960 (DCT) is
abnormal.
Serial communication with SY-961 (DCS) is
abnormal.
In measurement condition, induced voltage
exceeds 200[V].
In measurement condition, exciting current
exceeds 6[A].
It is forecast that the output voltage of the third
or fourth turns of the measurement yoke
exceeds 200V when 2nd is measured by the
Single Seat Measurement.
It is a measurement condition for the increment
voltage to exceed 400V
p-p
measurement.
It is a measurement condition for the increment
current to exceed 12A
p-p
measurement.
OSC output is tried to be set under 0V during
approximate value excitation.
The exciting current is insufficient by asymptote
excitation.
OSC output is tried to be set under 0V during
asymptote excitation.
Induced voltage measured exceeds 200[V].
If Hm is specified in Fixed Parameter,
numeric values and units of Le and N
If Bm is specified in Fixed Parameter, confirm that
numeric numbers and units of Ae and N
correct.
Restart this product and SY-956.
When this phenomenon happens frequently,
contact Iwatsu service center to tell the error
message No.
Restart this product, SY-960 and SY-961.
When this phenomenon happens frequently,
contact Iwatsu service center to tell the error
message No.
Restart this product, SY-960 and SY-961.
When this phenomenon happens frequently,
contact Iwatsu service center to tell the error
message No.
Confirm that values and units of N2, Ae, freq. are proper.
Confirm that values and units of Bm specified in
Fixed Parameter is proper.
Reduce N2 of the sample.
Confirm that values and units of N
Confirm that value of Hm specified in Fixed
Parameter is proper.
Increase N
Confirm whether the value specified with Fixed
Parameter is too large.
Confirm whether the value of Ae is too large.
Confirm that values and units of N2, Ae, freq. are proper.
by the DC bias
Confirm that values and units of Bm specified in
Fixed Parameter is proper.
Reduce N2 of the sample.
Confirm that values and units of N
by the DC bias
Confirm that value of Hm specified in Fixed
Parameter is proper.
Increase N
Confirm whether the value specified with Fixed
Parameter is too small.
Reduce the gain of the power amplifier.
Confirm whether the sample is connected or
disconnected.
Confirm whether the output of the power amplifier
is proper.
Reduce the gain of the power amplifier.
Confirm that values and units of N
are proper.
Confirm that value and unit of Bm specified in
Fixed Parameter are proper.
Reduce N
51
B-H ANALYZER SY-8218 / SY-8219
Remedies
are correct.
1
and Le are proper.
1
of the sample.
1
and Le are proper.
1
of the sample.
1
, Ae, and freq.
2
of the sample.
2
confirm
that
are
2