5.4 Setting Measuring Method

● It sets the measurement method.
➀ Press
to display Configuration screen (seeFig.5-2).
➁ Move the edition cursor to List Box of a measurement item for which
Measuring Method is to be changed and press
➂ When Popup List opens, move the edition cursor to the item to be set and press
to determine the selection.
Table 5-4 shows description of each item and selectable items.
➃ Press
or
If another item needs to be changed, Configuration screen
not be closed.
need
[Explanatory note] Bold: Factory settings
Items
Average
Retry
Maint Retry
Mov-Avg
Coil
5-2: Increased number of times for averaging allows higher quality measurement. However, if the number of times for averaging is set too
large, the measurement time may become long or temperature of a measured sample may rise depending on measured samples or
measurement conditions, resulting in effect on a measurement value. Care should be taken.
5-3: Normally, 8 are recommended for a default number of times. If the number is too large, the number of times for excitation increases
which causes temperature of a sample to rise, resulting in damage to the sample.
5-4: Increased number of times for moving average allows noise ingredients included in measurement signals to be reduced. However, if
the number is too large, the original measurement signal may deteriorate.
to close Configuration screen.
Table 5-4 Setting of measurement method
Selectable items
0
6
Power of 2
- 2
Selects No. of averaging times of measurement signal waveform
16
64)
3
5
Power of 2
- 2
8
Selects the upper limited No. of times of excitation until values of specified Fixed
3
5
Power of 2
- 2
Parameter (target parameter) converge within Tolerance
Free
Maint Retry mode is used only for service and not disclosed to a customer.
8
Odd in 1 - 111
Selects the moving average number applied to sample data of a measurement signal
waveform
3
1
2
Measuring Method
.
*5-4
.
1-coil method
Selects
2-coil method
measure a sample.
28
B-H ANALYZER SY-8218 / SY-8219
Setting of Equipment
Fig.5-2 Configuration screen
Description
*5-2
*5-3
1-coil method
2-coil method
or
(1, 2, 4, 8, 16, 32,