Iwatsu SY-8218 Instruction Manual page 54

Table of Contents

Advertisement

Scaled-up B-H graph
● The time axis graph is scaled up/ down.
① Press
to display the time axis graph (see Fig.6-12).
② When B scale is to be scaled, move the edition cursor to the B scale List Box; when H scale is to be scaled, to the H scale List Box; when V
scale is to be scaled, to the V scale List Box; or when I scale is to be scaled, to the I scale List Box.
And then press
➂ When Popup List opens, move the edition cursor to the scale to be set and press
If a value larger than a current setting is selected, the displayed graph is scaled down, and if a smaller value is selected, the displayed graph
is scaled up (see Fig.6-13).
④ Scales to be set are shown in Table 6-6. Selection of RESET allows the displayed graph to return the original scale.
Items
Unit
H scale
[A/m]/div
B scale
[mT]/div
I scale
[A]/div
V scale
[V]/div
Fig.6-11 Scaling up of Fig.6-10
. The time axis cannot be scaled up/ down.
Table 6-6 Selectable scales
Reset, 0.01, 0.02, 0.05, 0.1, 0.2, 0.5, 1, 2, 5, 10, 20, 50, 100, 200, 500, 1000, 2000, 5000, 10000
Reset, 0.01, 0.02, 0.05, 0.1, 0.2, 0.5, 1, 2, 5, 10, 20, 50, 100, 200, 500, 1000
Reset, 1E-6, 2E-6, 5E-6, 10E-6, 20E-6, 50E-6, 100E-6, 200E-6, 500E-6, 0.001, 0.002, 0.005, 0.01, 0.02, 0.05,
0.1, 0.2, 0.5, 1, 2, 5
Reset, 1E-6, 2E-6, 5E-6, 10E-6, 20E-6, 50E-6, 100E-6, 200E-6, 500E-6, 0.001, 0.002, 0.005, 0.01, 0.02, 0.05,
0.1, 0.2, 0.5, 1, 2, 5, 10, 20, 50, 100
to determine the scale.
Selectable scales
42
B-H ANALYZER SY-8218 / SY-8219