Iwatsu HV-P60A Instruction Manual

High voltage probe

Advertisement

Quick Links

Instruction Manual
High Voltage Probe
HV-P60A

Advertisement

Table of Contents
loading

Summary of Contents for Iwatsu HV-P60A

  • Page 1 Instruction Manual High Voltage Probe HV-P60A...
  • Page 2 Ⓒ2015 IWATSU TEST INSTRUMENTS CORPORATION.All rights reserved....
  • Page 3: Introduction

    High Voltage Probe HV-P60A Introduction ◇ Thank you for purchasing the Iwatsu's instrument. We hope that it can be used for a prolonged period. ◇ We would ask you to thoroughly read this instruction manual before commencing the operation and keep it at a readily accessible location for future reference.
  • Page 4 High Voltage Probe HV-P60A Read the following safety information Warnings ● Measure in accordance with operating procedure when using this instrument. Erroneous operating procedure (refer to “3. Adjustment of Probe” and “4. Measuring method”.) may result in an electric shock, fire or malfunction.
  • Page 5 Iwatsu office or our sales distributors for repair. The customer should never repair the instrument because it is dangerous.
  • Page 6 If not, an electric shock, fire or malfunction may result. If this instrument is wet with water or medcine, contact Iwatsu office or our sales distributors for repair. ● Do not directly hold this instrument when using If the instrument is directly held by the hand, never hold it because it may result in an electric shock.
  • Page 7 High Voltage Probe HV-P60A Read the following safety information Warnings ● Connect the playground of the probe body with the grounding potential (playground) of the object to be measured. When the playground of the probe body is connected with potential other...
  • Page 8 If you do not have a proper packing or shock-absorbing material for transportation, contact Iwatsu office or our sales distributors. If transportation is to be handled by a forwarding agent, display "Contains a precision machine" on each side of the packing box.
  • Page 9: Checking Packed Materials

    Checking packed materials When receiving this instrument, check the packed materials while referring to the following “Components.” If there is a missing item or an item damaged during transportation, immediately contact Iwatsu office or our sales distributors. Components ・Probe body ・・・・・・・・・1...
  • Page 10: Management Of Instrument

    Repair and sending instrument to be repaired If a malfunction occurs, send this instrument to Iwatsu office or our sales distributors. Malfunctions that occur during the warranty period due to our responsibility will be repaired by us free of charge.
  • Page 11: Table Of Contents

    High Voltage Probe HV-P60A Contents Introduction ..................i To ensure the Safety Operation ............i Checking packed materials .............. vii Components ..................vii Management of instrument ............. viii Repair and sending instrument to be repaired ......viii Cleaning of this instrument ............viii Storage .....................
  • Page 12: General Description

    High Voltage Probe HV-P60A 1. General Description This instrument is a passive prove to be connected to an oscilloscope to measure a high-voltage waveform. Its frequency bandwidth is DC to 50 MHz, maximum input voltage is 60 Kv, and attenation ratio is 2000:1.
  • Page 13: Carrying Bag (Accessory)

    High Voltage Probe HV-P60A 2.2 Carrying bag (accessory) A probe body and various accessories are bundled to Carrying bag like Figure 2, and delivered. Moreover, the probe body and the accessory can be carried by putting it in Carrying bag after packing things are opened.
  • Page 14: Adjustment Of Probe

    High Voltage Probe HV-P60A 3. Adjustment of Probe A waveform adjustment is necessary for this probe. A waveform adjustment of this probe adjusts the capacity of the variable capacitor that exists in this probe as the gain steadies for the frequency when the oscilloscope is combined with this probe.
  • Page 15: Adjustment

    High Voltage Probe HV-P60A 3.2 Adjustment This instrument is connected and adjusted according to the following procedures. (1) As shown in Figure 3(a), the ground terminal of the probe body is connected with the terminal GND of the signal generator.
  • Page 16 High Voltage Probe HV-P60A (4) Adjust the input range of the oscilloscope, and display the waveform on the screen of the oscilloscope. (5) The driver for the adjustment is inserted in the adjustment hole of the Matching box (Refer to Figure 3(b)). The waveform of the probe is adjusted to become the ideal waveform like “Figure 3(e)
  • Page 17: Measuring Method

    Probe body HV-P60A GND Terminal of Probe receiving part Probe stand Fixed screws Probe stand Hole plug SK-301 * Hole to pass cable from output of the bottom of Probe body Figure 4(a) Probe body HV-P60A and Probe stand SK-301...
  • Page 18 Figure 4(b) Probe stand SK-301 and Probe receiving part Note When buying it, probe stand SK-301 has installed the probe receiving part for HV-P30A. Replace the probe receiving part for HV-P60A and replace the top and bottom. (Refer to the manual of SK-301 for details.)
  • Page 19 High Voltage Probe HV-P60A Screw hole position for installation of Probe Probe receiving part (back) body (four places) Figure 4(c) Screw hole position for installation of Probe body on Probe receiving part...
  • Page 20 High Voltage Probe HV-P60A (2) Connection of oscilloscope and object to be measured (Refer to Figure 5.) 1) The power switch of the object to be measured is turned off. Caution! Connect following 2), 3) and 5) with each terminal firmly so that there is no shake.
  • Page 21 High Voltage Probe HV-P60A 5) The earth lead of the attachment is connected with the ground terminal of the object to be measured. 6) The lead wire is drawn out from the point to be measured, and connected with Tip of Probe. (Refer to Figure 6.) When connecting the lead wire, it is tightened firmly with Clamp of the attachment.
  • Page 22: Measuring Procedure

    High Voltage Probe HV-P60A 4.2 Measuring Procedure (1) Confirm that the voltage of the object to be measured is within the range of the measurement of this instrument. (2) Confirm that there is neither person in surroundings nor something that touches.
  • Page 23: Specifications

    High Voltage Probe HV-P60A 5. Specifications 5.1 Instrument specifications Maximum input voltage 60 kV Continuous wave (sine wave) Refer to "Figure 7 Maximum input voltage and frequency" on next page. 80 KV (Standard pulse wave: 1×40 μs) Single-shot pulse Frequency bandwidth DC to 50 MHz(-3dB)...
  • Page 24 High Voltage Probe HV-P60A Maximum input voltage (DC + Peak AC)
  • Page 25: Appearance

    High Voltage Probe HV-P60A 5.2 Appearance Unit: mm...
  • Page 26 Address : 7-41 Kugayama 1-chome Suginami-ku Tokyo, 168-8511 Japan Phone : +81 3 5370 5483 Facsimile : +81 3 5370 5492 Homepage : http://www.iti.iwatsu.co.jp...
  • Page 27 HV-P60A...