Instruction Manual Rogowski Coil Current Probe SS-281A(-H)/282A(-H) SS-283A(-H)/284A(-H) SS-285A(-H)/286A(-H) SS-287A(-H)/288A(-H) SS-289A(-H) Distributed by: Sie haben Fragen oder wünschen eine Beratung? Angebotsanfrage unter +49 7121 / 51 50 50 oder über info@datatec.eu...
Page 2
Ⓒ2015 IWATSU ELECTRIC CO., LTD.All rights reserved....
◇ Reproduction or reprinting of the contents of this manual without prior permission from IWATSU is prohibited. ◇ If any question arises about this instrument, contact IWATSU or our sales distributors. Revision History ◇ March 2015: 1st edition ◇...
(remove the AC adaptor from the outlet when used). Continued use under these circumstances may result in an electric shock or fire. Contact IWATSU or our sales distributors for repair. Do not attempt to repair this instrument yourself. Never use this instrument for measurement where current or frequency to be measured exceeds the specifications (refer to the absolute maximum di/dt in Section 5.1 of this manual).
Page 7
BNC cable or AC adaptor (option) of this instrument. If not, an electric shock, fire, or failure may occur. Do not use any of the cables or AC adaptor if damaged. Contact IWATSU or our sales distributors for repair.
Page 8
Do not use this instrument when it has failed. Using a failed main unit, cable or AC adaptor may be a cause of an electric shock or fire. In the event of a failure, contact IWATSU or our sales distributors for repair.
Read the following safety information. (continued) CAUTION Hold the plug when removing the AC adaptor from the outlet. Pulling on the AC adaptor cable will damage the cable. This may be a cause of an electric shock, fire, or failure. ...
Page 10
In addition, wipe off with a dry cloth when water droplets are attached. If not, an electric shock or failure may occur. We recommend that you contact IWATSU or our sales distributors for periodic inspection and calibration about once a year.
Verify packed Items When receiving this instrument, verify the packed items while referring to the following “Components.” If there is a missing item or an item damaged during transportation, immediately contact IWATSU or our sales distributors. Components ・・・・・1 Rogowski Coil Current Probe ...
Repair and sending instrument to be repaired If a failure occurs, send the instrument to IWATSU or our sales distributors. When sending an instrument to be repaired, clearly write the instrument...
Chapter 1 Overview Chapter 1 Overview 1.1 Rogowski Coil Current Probe This instrument is a current probe that uses the Rogowski Coil as a sensor. The electric current of the place etc. where the terminal and the electric circuit of the semiconductor device are complicated can be measured by using this instrument.
Chapter 1 Overview 1.2 Features The following describes the features of this instrument. Targets of measurement • Semiconductor switch waveforms • Transient current and pulsed current etc. of electrical instrument Compact measurement The Rogowski Coil that detects the electric current thinly has the following advantages for the removable type (In the length of the coil, it is 84mm, and the diameter in the coil section is 1.7mm or less).
Chapter 1 Overview 1.3 Usage Example To measure current using this current probe, place a conductor to be measured through the coil loop, and detect a change in magnetic flux caused by the current flowing through the conductor with the Rogowski Coil as an electromotive force, as shown in the following usage example (conceptual illustration).
Chapter 2 Names and Functions of Parts 2.2 Names and Functions of Main Unit Name Description Reference POWER switch ► Section 4.2 Turns this instrument on and off. This instrument is turned on when this slide switch is moved to the I side and is turned off when moved to the O side.
Chapter 2 Names and Functions of Parts 2.3 Rogowski Coil Sensor ② ④ ① Current direction mark ③ Name Description Rogowski Coil A coil for detecting the current in conductors to be measured. Forms a closed loop during measurement. Covered with an insulating outer sheath for heat resistance and insulation.
Chapter 3 Measurement Principle Chapter 3 Measurement Principle This instrument is designed to measure the current (current waveform) in conductors to be measured. It consists of the Rogowski Coil Sensor and the main unit (integrator). Sections 3.1 and 3.2 in this chapter describe measurement principles and functions of each part and are concluded with an explanation of a process of measuring the current (current waveform).
Chapter 3 Measurement Principle 3.1 Rogowski Coil Sensor Figure 3.1 shows that the current I flows through the conductor to be measured. This section describes the principle of detecting this current by the Rogowski Coil step by step. Signal that is Output voltage (V) differential of input (Current waveform)
Chapter 3 Measurement Principle 3.2 Main Unit As illustrated in Figure 3.2, the electromotive force e induced in Section 3.1 is brought in the input part of the main unit through the sensor cable. Observation with an oscilloscope (not included) Main unit (in the dotted area) and other monitoring devices...
Chapter 4 Measurement Chapter 4 Measurement 4.1 Preparation for Measurement (1) Have this instrument, a monitoring device, and a target to be measured ready. * A monitoring device means an oscilloscope, DVM, recorder, and other devices. (2) Place four AA dry batteries in the main unit, or connect the AC adaptor (option) to this instrument.
Chapter 4 Measurement 4.2 Measurement Procedure (1) Make sure that all preparations described in the previous section are completed. (2) Set up a target to be measured so that it can be used. Do not energize the target to be measured even when it is ready to be used.
Page 24
Chapter 4 Measurement The current detection polarity is shown as in the following figure. Positive polarity is output by current flowing in the direction indicated by the current direction mark shown on the joint box. Current to be measured Figure 4.2 Current Detection Polarity (4) Turn on this instrument (this instrument is turned on by turning the slide switch shown in the following figure to the I side and off by moving the switch to the O side).
Page 25
Chapter 4 Measurement ・ When adjusting the DC offset voltage (zero CAUTION point), included accessory Adjustment Screwdriver, align the tip of the screwdriver with the adjustment slot, and turn it without applying excessive pressure. The required rotation torque is less than 20 mN・m If using a screwdriver other than the accessories or applying excessive pressure to ZERO ADJ of the main unit while you don’t align the included accessory...
Chapter 4 Measurement 4.3 Before Measurement This section describes notes on accuracy and the handling of the Rogowski Coil Sensor when measuring the current using this instrument. 4.3.1 To Achieve Accurate Measurement As explained in Chapter 3 “Measurement Principle,” the Rogowski Coil is a wire rod wound tight around a tube-like wick material, as illustrated in Figure 3.1.
Page 27
Chapter 4 Measurement 4.3.2 To Minimize Influence of Fluctuations in Voltage If there is a part whose voltage fluctuates around the Rogowski Coil Sensor, electrostatic coupling between the part and the Rogowski Coil will occur. This may influence the output from this instrument. If possible, place the Rogowski Coil Sensor away from the part whose voltage fluctuates.
Page 28
We recommend that you replace a deformed coil or a coil with its surface damaged, in order to maintain safety and accurate measurement. Contact Iwatsu office or our sales distributors for replacement.
Page 34
SS-281A(-H)/282A(-H)/283A(-H)/284A(-H)/285A(-H) SS-286A(-H)/287A(-H)/288A(-H)/289A(-H) Distributed by: Sie haben Fragen oder wünschen eine Beratung? Angebotsanfrage unter +49 7121 / 51 50 50 oder über info@datatec.eu...