Vacuum Products Division

1.2.1 Solvents

Warning

1.2.2 Vacuum Equipment and Cleanliness

Caution
Manual 699910005, Hot Ion Source Replacement
The mechanical components of leak detectors may be cleaned with one
of the recommended solvents. When heated, sprayed, or exposed to
high-temperature equipment, these solvents become flammable and
explosive, causing serious injury or death. Do not use these solvents near
a high-temperature source. Ventilate the working area with a blower and
work in a large, well-ventilated room.
Solvents are irritants, narcotics, depressants and/or carcinogens. Their
inhalation and/or ingestion may produce serious side effects. Prolonged
or continued contact with the skin results in absorption through the skin
and moderate toxicity. Always ensure that cleaning operations are
carried out in large, well ventilated rooms, and wear eye shields, gloves,
and protective clothing.
Due to the effective cleaning nature of VacuSolv solvent and its residue-
free properties, Agilent' Component and Spectrometer Cleaning Kit (Part
Number 670029096), used in accordance with the kit instructions, is
recommended for cleaning spectrometer components. The kit can also
be used for fine cleaning of other parts in the leak detector's vacuum
system such as valves and fittings. No rinsing steps or high-temperature
drying is required following cleaning with VacuSolv. Although appropriate
precautions are advised, VacuSolv is compatible with most materials and
does not contain toxic chemicals or CFCs (chlorofluorocarbons). Other
acceptable solvents are isopropyl alcohol (IPA) or Dow Corning
To clean the leak detector plastic top enclosure, the LCD display and
front panels, use only a soft cloth slightly dampened with water or a mild
soap.
Do NOT use excess water or cleaning solvents of any kind.
Avoid splashing any cleaning solvents into the unit through the
ventilation openings or Front Panel buttons. Wipe the surface with a dry
lint-free cloth.
Wear non-powdered, ESD-safe Nitride or equivalent gloves to prevent
skin oils from getting on spectrometer internal components.
OS-20.
®