Download Print this page

Agilent Technologies VS C15 Series Field Replacement Instructions

Calibrated leak replacement for leak detectors

Advertisement

Quick Links

Calibrated Leak
Replacement for VS
C15, VS, and
(G8610, G8611, or
G8612) Series Leak
Detectors
Part Number 699909998
Rev. C
July 2017
Vacuum Products Division
FIELD REPLACEMENT
INSTRUCTIONS

Advertisement

loading

Summary of Contents for Agilent Technologies VS C15 Series

  • Page 1 Vacuum Products Division Calibrated Leak FIELD REPLACEMENT INSTRUCTIONS Replacement for VS C15, VS, and (G8610, G8611, or G8612) Series Leak Detectors Part Number 699909998 Rev. C July 2017...
  • Page 2 Calibrated Leak Replacement for VS C15, VS, and (G8610, G8611, or G8612) Series Leak Detectors This page intentionally left blank.
  • Page 3 Agilent Field Instruction Sheet Calibrated Leak Replacement for VS C15, VS, and (G8610, G8611, or G8612) Series Leak Detectors Preface This instruction sheet gives the procedure for replacing the calibrated leak for both the VS C15 (“Section One: VS C15 Calibrated Leak Replacement”...
  • Page 4 Solvents are irritants, narcotics, depressants and/or carcinogens. Their inhalation and/or ingestion may produce serious side effects. Prolonged or continued contact with the skin results in absorption through the skin and moderate toxicity. Always ensure that cleaning operations are carried out in large, well-ventilated rooms, and wear eye shields, gloves, and protective clothing.
  • Page 5: Section One: Vs C15 Calibrated Leak Replacement

    Section One: VS C15 Calibrated Leak Replacement Equipment Required • M4 Allen Wrench • Flat Head Screwdriver Installation Procedure WARNING Disconnect power from the unit before performing any maintenance procedure that requires physically disconnecting any part of the system. 1. Disconnect the power to back of the unit and unplug. 2.
  • Page 6 4. Loosen and remove the clamp (Figure 2: Remove Clamp). Figure 2: Remove Clamp 5. Disconnect the cable from the calibrated leak board (Figure 3: Cable to Calibrated Leak Connection). Figure 3: Cable to Calibrated Leak Connection Calibrated Leak Replacement for VS C15, VS, and (G8610, G8611, or G8612) Series Leak Detectors Page 4 of 12 Document No.
  • Page 7 6. Remove the assembly and use a flat head screw driver to remove the two screws holding the calibrated leak (Figure 4: Calibrated Leak Screws). Figure 4: Calibrated Leak Screws 7. Discard the calibrated leak and the O-ring. 8. Inspect the replacement calibrated leak and O-ring for damage or particle contamination and remove any particle contamination.
  • Page 8 11. Reattach the leak assembly to the flange with the clamp. 12. Close the cover and secure using the existing hardware. 13. Power up the VS C15 component leak detector. 14. Set up the Internal Calibrated Leak by either: • Setting up the Internal Calibrated Leak Option via the front panel display by: a.
  • Page 9: Section Two: Calibrated Leak Replacement For Vs Series Leak Detectors

    Section Two: Calibrated Leak Replacement for VS Series Leak Detectors Equipment Required • Extended Length M5 Allen Wrench (included with leak detector) • Slotted Screw Driver Installation Procedure For clarity, some items have been omitted from views. Figure 1: Rear Cover Screws WARNING Disconnect power from the unit before performing any maintenance procedure that requires physically disconnecting any part of the system.
  • Page 10 Figure 2: Calibrated Leak Location 4. Carefully disconnect the calibrated leak temperature cable from the calibrated leak PC board. 5. Using a slotted screwdriver, remove the two screws holding the calibrated leak in place (Figure 2: Calibrated Leak Location). 6. Discard the calibrated leak and the O-ring. 7.
  • Page 11 WARNING Extreme care must be taken due to the presence of high voltage. Ensure that high voltage shield and front cover are in position before proceeding (Figure 3: High Voltage Location). 10. Connect the power cord and power up the unit. 11.
  • Page 12: Section Three: Calibrated Leak Replacement For G8610, G8611, Or G8612

    Section Three: Calibrated Leak Replacement for G8610, G8611, or G8612 Equipment Required • Slotted Screw Driver Installation Procedure For clarity, some items have been omitted from views. Figure 1: Rear Cover Screws WARNING Disconnect power from the unit before performing any maintenance procedure that requires physically disconnecting any part of the system.
  • Page 13 Figure 2: Calibrated Leak Location 4. Carefully disconnect the calibrated leak temperature cable from the calibrated leak PC board. 5. Using a slotted screwdriver, remove the two screws holding the calibrated leak in place (Figure 2: Calibrated Leak Location). 6. Discard the calibrated leak and the O-ring. 7.
  • Page 14 WARNING Extreme care must be taken due to the presence of high voltage. Ensure that high voltage shield and front cover are in position before proceeding (Figure 3: High Voltage Location). 10. Connect the power cord and power up the unit. 11.
  • Page 15 Please follow these instructions whenever one of our products needs to be returned. Complete the attached Request for Return form and send it to Agilent Technologies (see below), taking particular care to include the completed Health and Safety declaration Section. No work can be started on your unit until we receive a completed copy of this form.
  • Page 16: Terms And Conditions

    Terms and conditions TERMS AND CONDITIONS Please read the terms and conditions below as they apply to all returns and are in addition to the Agilent Technologies Vacuum Product Division – Products and Services Terms of Sale. Unless otherwise pre-negotiated, customer is responsible for the freight charges for the returning product. Return shipments must comply with all applicable Shipping Regulations (IATA, DOT, etc.) and carrier requirements.
  • Page 17 Vacuum Products Division Request for Return Form Customer information Company : Contact Name: Address: Tel: Fax: Email: Equipment Product description Agilent PartNo Agilent Serial No Original Purchasing Reference Failure description Type of process (for which the equipment was used) Type of return Non Billable Billable New PO # (hard copy must be submitted with this form): ______________________________________...
  • Page 18 Agilent Vacuum Products Division/Sales and Service Offices United States India (Service) Southeast Asia Agilent Technologies Agilent Technologies India Pvt. Ltd. Agilent Technologies Sales Sdn Bhd 121 Hartwell Avenue C-Block, RMZ Centennial Plot Number -8A, 8B, Unit 201, Level 2 uptown 2,...

This manual is also suitable for:

Vs seriesG8610 seriesG8611 seriesG8612 series