17. Connect the collimator wire to the collimator plate using an M 1.5 screw. See Figure 2-6.
18. Connect the 4 coupler screws to their respective filaments.
19. Insert the high voltage ion source wire (brownish-orange wire in rendering) into the ion
chamber and tighten the M1.6 socket head screw.
NOTE
Manual 699910005, Hot Ion Source Replacement
Collimator
screw
Figure 2-6. Collimator wire connection
Pull lightly with small pliers on the high voltage ion source wire
(brownish-orange) to ensure it is connected securely and will not pull out.
Figure 2-7. Ion chamber and extractor plate
Extractor
plate
Ion
chamber