Agilent Technologies Helium G8610 Series Field Installation Instructions page 20

Ion source replacement for leak detection
Hide thumbs Also See for Helium G8610 Series:
Table of Contents

Advertisement

Vacuum Products Division
28. On the PCB, use the resistance meter to verify an open circuit between any two of the six
ion source header pins (except FIL-1 to FIL-1 and FIL-2 to FIL-2 which should be 0.3 Ohms
or less). Also verify an open circuit between the body of the spectrometer and any of the
ion source header pins (Error! Reference source not found.
circuit) at any of the points, remove the header and inspect for shorting.
29. If successful, leak check the spectrometer to ensure a leak free seal.
30. Place the spectrometer cover over the header and align the holes (Figure 2-13Error!
Reference source not found.).
Reference source not found.
Reference source not found.
Reference source not found.
31. Install screws and washers (items 5 & 6), then tighten using a M3 Phillips screw driver
(Error! Reference source not found.
Error! Reference source not found.
Error! Reference source not found.
Error! Reference source not found.). Tighten one screw over the ground strap connector.
32. Re-plug in both the pre-amp and ion source cable.
33. Reinstall the leak detector covers and fasten hardware.
34. Power up the leak detector.
35. Let the unit run for 20-30 minutes, then perform a calibration (via the I/O, front panel
display, or RS232), per the user manual, to validate a successful installation.
Manual 699910005, Hot Ion Source Replacement
Figure 2-12. Ion source header pin schematic
Error! Reference source not found.). If there is continuity (short
Error! Reference source not found.
Error! Reference source not found.
Figure 2-13. Ion source header pin locations
Error!
Error!
Error!
Page 15

Advertisement

Table of Contents
loading

This manual is also suitable for:

Helium g8611 seriesHelium g8612 seriesHelium vs series

Table of Contents