Agilent Technologies Helium G8610 Series Field Installation Instructions page 31

Ion source replacement for leak detection
Hide thumbs Also See for Helium G8610 Series:
Table of Contents

Advertisement

Vacuum Products Division
30. On the PCB, use the resistance meter to verify an open circuit between any two of the six ion
source header pins (except FIL-1 to FIL-1 and FIL-2 to FIL-2 which should be 0.3 Ohms or less).
Also verify an open circuit between the body of the spectrometer and any of the ion source header
pins (Figure 3-14). If there is continuity (short circuit) at any of the points, remove the header and
inspect for shorting.
31. If successful, leak check the spectrometer to ensure a leak free seal.
32. Place the spectrometer cover over the header and align the holes (Figure 2-13Figure 3-15).
33. Install screws and washers (items 5 & 6), then tighten using a M3 Phillips screw driver (Error!
Reference source not found.). Tighten one screw over the ground strap connector.
Reference source not found.
Reference source not found.
Reference source not found.
34. Re-plug in both the pre-amp and ion source cable.
35. Reinstall the leak detector covers and fasten hardware.
36. Power up the leak detector.
37. Let the unit run for 20-30 minutes, then perform a calibration (via the I/O, front panel display, or
RS232), per the user manual, to validate a successful installation.
Manual 699910005, Hot Ion Source Replacement
Figure 3-14. Ion source header pin schematic
Figure 3-15. Ion source header pin locations
Error!
Error!
Error!
Page 26