Agilent Technologies Helium G8610 Series Field Installation Instructions page 27

Ion source replacement for leak detection
Hide thumbs Also See for Helium G8610 Series:
Table of Contents

Advertisement

Vacuum Products Division
12. Loosen the four filament coupler screws.
13. Loosen the collimator screws.
14. Loosen two M1.6 socket head screws securing the ion source high voltage connection.
15. On a clean work surface, remove the two M4 support screws. See Figure 3-7.
Collimator
16. Gently remove the ion support assembly from the spectrometer header.
NOTE
17. Position the new ion support assembly into the spectrometer header and align.
18. Secure the spectrometer header using two M4 support screws
Manual 699910005, Hot Ion Source Replacement
screw
Figure 3-7. M4 support screws
Ensure that no damage is done to the spectrometer wire during
disassembly.
M4
Support
Screws
Page 22

Advertisement

Table of Contents
loading

This manual is also suitable for:

Helium g8611 seriesHelium g8612 seriesHelium vs series

Table of Contents