Spectrometer - Agilent Technologies Helium G8610 Series Field Installation Instructions

Ion source replacement for leak detection
Hide thumbs Also See for Helium G8610 Series:
Table of Contents

Advertisement

Vacuum Products Division

1.2.5 Spectrometer

CAUTION
CAUTION
Manual 699910005, Hot Ion Source Replacement
Store the Ion Source/Preamplifier sub-assembly in a cool, dry area in a
tightly sealed, ESD protected container. Wear non-powdered, ESD-safe
Nitride or equivalent gloves when handling the spectrometer. Wash
hands thoroughly after handling the spectrometer filaments and
especially before smoking or eating.
The spectrometer and PCB's are static sensitive devices. Wear a
grounding strap when performing any maintenance on these units and
especially when performing maintenance of static sensitive parts.
The spectrometer operates at a very high vacuum produced by the high
vacuum turbomolecular pump. Service of the spectrometer requires that
this vacuum be vented to the atmosphere.
Page 5

Advertisement

Table of Contents
loading

This manual is also suitable for:

Helium g8611 seriesHelium g8612 seriesHelium vs series